BD Biosciences™ Setup Beads (CE/IVD)
Description
- Set voltages appropriate for the BD multicolor tube assay.
- Fluorescent particles that are used to set cytometer detector photomultiplier tube voltages (PMTVs) for the BD multicolor tube assay.
- PMTVs are manually adjusted to place the BD OneFlow Setup beads at their lot specific median fluorescence intensity (MFI) target ranges for all fluorescence parameters.
- Lysed washed blood (LWB) is used to set cytometer FSC and SSC voltages to a target value range.
- The detector settings are then saved as Application Settings.
Specifications
Specifications
| Content And Storage | One vial of BD OneFlow Setup beads, Monthly MFI target range card, Daily MFI target range card |
| No. of Tests | 25 Tests |
| Quantity | 25 Tests |
| Stability | After dilution, the beads are stable for 1 hour at 18%deg;C to 25°C, and for 8 hours at 2°C to 8°C |
| Product Type | Fluorescent Polystyrene Beads |