Thermo Scientific™ Eutech™ CON 150 Conductivity Meter
Description
Eutech™ CON 150 is an IP67-rated waterproof conductivity meter. View conductivity or TDS measurement with temperature as well as cell efficiency on a large three line LCD screen.
Features:
- Built-in memory function – Stores up to 150 measurements along with temperature
- Adjustable temperature coefficient – Set to 0.0 % to meet USP methods for non-compensated measurements
- Power with AA batteries or use optional AC adapter
- Calibrate using automatically recognized standards or manually; to a single point across all ranges or to multiple points (one calibration point per each conductivity range)
- Calibration due alarm – Set a visible reminder when it is time to recalibrate the meter
- Accepts probes with a 0.1, 1.0, or 10 cell constant
- Adjustable conductivity-to-TDS conversion factor – From 0.4 to 1.0
- Selectable manual or automatic temperature compensation (ATC) – For the highest accuracy in any situation
- Selectable ºF/ºC – Choose units to best fit your application
- Ready indicator, manual or automatic measurement hold feature and optional automatic shutoff function
- Includes our easy to clean conductivity cell
- Built-in stand – For easy benchtop or long-term use
- Wall-mount option – Ideal to save bench space
- Innovative Grip-Clip™ – Quickly and easily secures the probe to your beaker or container
Spécification
Spécification
| Description | CON 150 Meter Kit:Meter, conductivity cell (K = 1.0), Grip-Clip™, two AA batteries, hard carry case, calibration solution |
| Accuracy (Conductivity) | ±1% Full-scale |
| Measurement Modes | Conductivity, TDS with Temperature |
| Compatible Probes | Probes with a 0.1, 1.0, or 10 cell constant |
| Calibration | TDS: Manual 1 point/range; Temperature: Off set in 0.1 increments, ±10 ºC or ±18 |
| Calibration Points (Conductivity) | Auto/manual; 1 point/range |
| Range (Conductivity) | 0 to 200 mS |
| Resolution (Conductivity) | 0.01 μS to 0.1 mS |
| Temperature Compensation | Automatic or Manual (-10 to 110°C) |
| Range (TDS) | 0 to 200 ppt (TDS factor 1.0) |
| Afficher plus |
Nom du produit
En cliquant sur Soumettre, vous reconnaissez que vous pouvez être contacté par Fisher Scientific au sujet des informations que vous avez fournies dans ce formulaire. Nous ne partagerons pas vos informations à d'autres fins. Toutes les informations de contact fournies seront également conservées conformément à notre politique de confidentialité. Politique de confidentialité.