missing translation for 'onlineSavingsMsg'
Learn More

Thermo Scientific™ Evolution™ 260 Bio UV-Visible Spectrophotometer

The Thermo Scientific™ Evolution 260 spectrophotometer offers reliable, high-quality performance and enhanced customer usability.

Brand:  Thermo Scientific™ 840-211000


Additional Details : Weight : 0.50000kg

 View more versions of this product

Product Code. 10438278

This item is currently unavailable or has been discontinued.
View the product page for possible alternatives.
View alternative products »


Description

Description

New sales of this model have been discontinued. Please see the newest models: the Evolution Pro Spectrophotometers.

Complete Life Science Applications

Next generation INSIGHT software provides the streamlined convenience of easy-to-understand nucleic acid and protein analysis methods and includes dedicated modules for:

  • Nucleic Acid concentration by direct UV measurement at 260nm including DNA, ssDNA, RNA, and siRNA
  • Nucleic acid purity measurements from 260/280nm and 260/230 ratios
  • Protein analysis by direct UV measurement at 205 or 280nm
  • Most common colorimetric analysis methods including: Pierce 660nm Protein assay, Bradford, BCA and Lowry
  • Measuring the fluorescence labeling efficiency for proteins and nucleic acids
  • Measuring the microarray labeling efficiency with one or two dyes
  • Customizable, Straightforward INSIGHT Software

 

Personalize your home screen for easy navigation and convenience

  • Workbook tools save time and ensure accurate results
  • CUE scripting allows creation of customized interface and simplified workflow

 

Best-in-Class Performance

  • Double-beam geometry ideal for kinetics or any sample that may change over time
  • Application Focused Beam Geometry (AFBG) technology provides superior results for microcells and turbid samples
  • Xenon flash lamp requires no warm-up time allowing instant measurements
  • Electronic triggering provides the highest level of accuracy for rapid-mixing kinetics measurements
  • Precision monochromator drive delivers fast scanning data collection and wavelength accuracy Powerful Convenience at your Fingertips
  • Color, touch screen provides instrument control using built-in computer
  • Integrated keypad allows communication with INSIGHT software
  • Choice of Local or PC configurations provides flexibility and convenience

 

Accessories for all your Sampling Needs

  • Wide accessory selection with various temperature control options, sample changers, sippers and solid sampling devices complete your analysis
  • Automated cell changers with Peltier temperature control increase sample throughput
  • Choice of Peltier Single Cell Holder or a Smart 8-Cell Peltier system for temperature control and sample monitoring from 0° to 100°C
  • Easy software integration gives you complete control of your measurements

 

Reliable Assurance of System Performance

  • The Thermo Scientific UV Validator package provides support for system qualification and validation activities
  • Complete documentation facilitates compliance of your system with the requirements of FDA, GxP, ISO 9001:2000 and ISPE 2001 guidelines and regulations
  • CVC and Mercury Lamp accessory automated testing ensures accuracy of your data
  • Thermo Scientific INSIGHTsecurity software provides complete tools to meet 21 CFR Part 11 requirements
Specifications

Specifications

±0.0010 A, 200 to 800 nm, 1.0 nm SBW, smoothing
Evolution 260 Bio UV-Vis Spectrophotometer, Computer Control, includes Euro/US/UK power cords
62.2 x 48.6 x 27.9 cm (24 x 19 x 11 in.)
Variable: 1.0 nm, 2.0 nm
190 to 1100 nm
100/240 V, 50/60 Hz selected automatically, 150 W maximum
7 years typical
AFBG microcell optimized, AFBG fiber optic optimized, AFBG materials optimized
-0.3 - 4.0A
±0.0002 A
Absorbance, % Transmittance, % Reflectance, Kubelka-Munk, log (1/R), log (Abs), Abs*Factor, Intensity
3 Years
10, 5, 2, 1.0, 0.5, 0.2, 0.1 nm
14.4 kg
ISO 9001:2008
Dual Silicon Photodiodes
Evolution 260 Bio
Computer Control UV-Vis Spectrophotometer
<0.0005 A/hr., 500 nm, 1.0 nm SBW, 1 hr. warmup
EU/US/UK power cords
0A: <0.00015A
1A: <0.00025A
2A: <0.00080A
260nm, 1.0nm SBW, RMS
(Guaranteed Performance Specifications) Resolution (toluene in Hexane): ≥1.8A Photometric Accuracy (60mg/L K2Cr2O7): ±0.010A Stray Light: ≤1%T at 198nm: KCI; ≤0.05%AT at 220nm: Nal, KI Wavelength Accuracy: ±0.5nm 541.9, 546.1nm Hg emission lines, ±0.8nm full range Wavelength Repeatability: ≤0.05nm, repetitive scanning of 546.1nm Hg emission line
>3.5A
Microsoft Windows 7, Windows 8
<1 to 6000 nm/min. (Variable)
±0.5 nm (541.9 nm mercury lines), ±0.8 nm (full range 190 to 1100 nm)
32 lb.